Chapter 16 453
GPIB Command Reference
THRR
16. GPIB
Command Reference
( S - Z , :PROG )
for the linear Y axis format.)
For the linear Y axis format
height is a threshold value of the distance between the maximal value
(or minimal value) and the measurement point value (in
Figure 16-9,
value of |A-A
L
|, |C-C
R
|, and so on).
For the log Y axis format
height is a threshold value of the ratio of the larger value to the smaller
value: maximal value (or minimal value) and measurement point value
(in
Figure 16-9, values of A/A
L
and C
R
/C, and so on). More
specifically, the threshold value shown below.
When H > 0, L > 0: H / L
When H < 0, L < 0: L / H (for negative values, a larger value has a
smaller absolute value, and therefore the denominator and
numerator are reversed.)
When H 0, L 0: No peak and negative peak are detected. Note
that, only when the value of heig ht is 1 (initial value), all maximal
values (or minimal values) are considered as peaks (or negative
peaks).
Where, H is the larger value and L is the smaller value. More
specifically, when judging whether a maximal value is a peak, H is the
maximal value and L is the measurement point value. When judging
whether a minimal value is a negative peak, H is the measurement
point value and L is the minimal value.
For the linear Y axis format, if height is set to 0 (initial value), all maximal values (or
minimal values) are considered as peaks (or negative peaks). On the other hand, for the log
Y axis format, if height is set to 1 (initial value), all maximal values (or minimal values)
are considered as peaks (or negative peaks).
Parameters
If the specified parameter does not reach the lower limit of the setting range, it is set to the
lower limit. If the specified parameter exceeds the upper limit of the setting range, an error
occurs and the command is ignored.
Query response {numeric}<newline><^END>
Corresponding
key
No front panel key is available to execute this function.
<numeric>
Description Condition of the peak/negative peak: height
Range For the linear Y axis format: 0 to 9.9E37
For the log Y axis format: 1 to 9.9E37
Initial value For the linear Y axis format: 0
For the log Y axis format: 1
Unit Varies depending on the measurement parameter (refer to the
explanation of unit in
“SCAL” on page 423).