Chapter 16 351
GPIB Command Reference
MKRAMPO
16. GPIB
Command Reference
( F - N )
MKRAMPO
Syntax MKRAMPO
Description For the active trace (set with the “TRAC” command on page 457), sets the measurement
parameter value at the marker position as the offset value of the limit for the measurement
parameter value. (No query)
Corresponding
key
[System] - LIMIT TEST - LIMIT LINE OFFSETS - MKR AMP. OFST.
MKRAUV?
Syntax MKRAUV?
Description For the active trace (set with the “TRAC” command on page 457), if the measurement
parameter is a vector value (for COMPLEX Z-Y), reads out the subsidiary readout of the
measurement parameter
value at the marker position. (Query only)
Query response {numeric}<newline><^END>
If the measurement parameter is a scalar value (for other than COMPLEX Z-Y), 0 is
always read out. If the measurement parameter is a vector value (for COMPLEX Z-Y), the
value is as follows.
If the polar chart format is selected, the phase is read out.
If the complex plane format is selected, reactance (for trace A) or susceptance (for trace
B) is reads out.
Corresponding
key
No front panel key is available to execute this function.
MKRCENT
Syntax MKRCENT
Description For the active trace (set with the “TRAC” command on page 457), sets the sweep
parameter value at the marker position as the center value of the sweep range. To set the
center value of each segment when creating the list sweep table, also use this command. If
the marker function is off, executing this command causes an error and the command is
ignored. (No query)
Corresponding
key
[Marker] - MKR CENTER