SPLD
16. GPIB
Command Reference
( S - Z , :PROG )
SPLD
Syntax SPLD {ON|OFF|1|0}
SPLD?
Description Enables/disables the vertically separate display for traces A and B.
If the separate display of traces A and B is enabled, the upper half of the LCD screen is
used as trace A measurement result display screen, and the lower half, as trace B
measurement result display screen. If the separate display of traces A and B is disabled, the
measurement results of traces A and B are superimposed on the same screen.
If the screen is set so that the upper half of the LCD screen is used as the measurement
result display screen and the lower half as the HP
Instrument BASIC screen (specified to
HIHB with the
“DISA” command on page 298), regardless of the setting to enable/disable
the separate display, both traces are superimposed on the same screen.
Parameters
Query response {1|0}<newline><^END>
Corresponding
key
[Display] - SPLIT on OFF
Description
ON or 1 Enables the separate display of traces A and B.
OFF or 0
(initial value)
Disables the separate display of traces A and B.