400 Chapter 16
GPIB Command Reference
PKDLTY
PKDLTY
Syntax PKDLTY <numeric>[OHM|DEG|RAD|SIE|H|F|PCT]
PKDLTY?
Description For the active trace (set with the “TRAC” command on page 457), sets the ΔY value (refer
to Figure 7-1 on page 97) parameter to define the peak searched for in the marker search
function.
Parameters
If the specified parameter is out of the allowable setting range, the minimum value (if the
lower limit of the range is not reached) or the maximum value (if the upper limit of the
range is exceeded) is set.
Query response {numeric}<newline><^END>
Corresponding
key
[Search] - PEAK - PEAK DEF MENU - PEAK DEF: ΔY
PKPOL
Syntax PKPOL {POS|NEG}
PKPOL?
Description For the active trace (set with the “TRAC” command on page 457), sets the polarity of the
peak searched for in the marker search function.
Parameters
Query response {POS|NEG}<newline><^END>
Corresponding
key
[Search] - PEAK - PEAK DEF MENU - PEAK PLRTY [ ]
<numeric>
Description ΔY value
Range 0 to 100E6
Initial value 1
Unit Varies depending on the measurement parameter (refer to the
explanation of unit in
“SCAL” on page 423).
Description
POS (initial value) Specifies positive peaks.
NEG Specifies negative peaks.