Chapter 16 287
GPIB Command Reference
DCALLOAD{R|L}
16. GPIB
Command Reference
( * , A - E )
DCALLOAD{R|L}
Syntax DCALLOADR <numeric>[OHM]
DCALLOADL <numeric>[H]
DCALLOAD{R|L}?
Description For user calibration data measurement (“CAL{A|B|C}” command on page 274), sets the
resistance value (the “DCALLOADR” command) and the inductance value (the
“DCALLOADL” command) of the LOAD standard of the calibration kit used to measure
LOAD calibration data.
Parameters
If the specified parameter is out of the allowable setting range, the minimum value (if the
lower limit of the range is not reached) or the maximum value (if the upper limit of the
range is exceeded) is set.
Query response {numeric}<newline><^END>
Corresponding
key
[CAL] - USER CAL -DEFINE VAUE - {LOAD RESIST (R) | LOAD INDUCT (L)}
<numeric>
Description Resistance value or inductance value
Range For resistance value: -100E9 to 100E9
For inductance value: -1E6 to 1E6
Initial value For resistance value: 50
For inductance value: 0
Unit For resistance value: Ω (ohm)
For inductance value: H (henry)
Resolution 1E-12