Agilent Technologies 4294A Service Manual Page 285

  • Download
  • Add to my manuals
  • Print
  • Page
    / 518
  • Table of contents
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 284
Chapter 16 285
GPIB Command Reference
CWFREQ
16. GPIB
Command Reference
( * , A - E )
CWFREQ
Syntax CWFREQ <numeric>[HZ|MHZ]
CWFREQ?
Description Sets the frequency of the oscillator for the oscillator (OSC) level sweep and dc bias level
sweep.
Parameters
If the specified parameter is out of the allowable setting range, the minimum value (if the
lower limit of the range is not reached) or the maximum value (if the upper limit of the
range is exceeded) is set.
Query response {numeric}<newline><^END>
Corresponding
key
[Source] - FREQUENCY
DATMEM
Syntax DATMEM
Description Copies the measured data into the memory array. It is copied to both the A and B
traces.(No query)
Corresponding
key
[Display] - DEF TRACR[ ] - DATA MEMORY
<numeric>
Description Frequency of the oscillator
Range 40 to 110E6
Initial value 1E6
Unit Hz
Resolution 1E-3
Page view 284