Agilent Technologies 4294A Service Manual Page 427

  • Download
  • Add to my manuals
  • Print
  • Page
    / 518
  • Table of contents
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 426
Chapter 16 427
GPIB Command Reference
SEAM
16. GPIB
Command Reference
( S - Z , :PROG )
SEAM
Syntax SEAM {MAX|MIN|TARG|PEAK|OFF}
SEAM?
Description For the active trace (set with the “TRAC” command on page 457), executes the specified
search function. If no search target is detected, the message saying “Not detected” is
displayed (no error occurs).
Parameters
Query response {MAX|MIN|TARG|PEAK|OFF}<newline><^END>
Corresponding
key
[Search] - {MAX | MIN | TRAGET | PEAK }
SEANPK
Syntax SEANPK
Description For the active trace (set with the “TRAC” command on page 457), moves the marker to the
next smaller peak of the last detected peak. If no peak is detected, the message saying “Not
detected” is displayed (no error occurs). If the marker function is off, executing this
command causes an error and the command is ignored. (No query)
Corresponding
key
[Search] - PEAK - NEXT PEAK
SEANPKL
Syntax SEANPKL
Description For the active trace (set with the “TRAC” command on page 457), moves the marker to a
peak to the left of the current marker position. If no peak is detected, the message saying
“Not detected” is displayed (no error occurs). If the marker function is off, executing this
command causes an error and the command is ignored. (No query)
Corresponding
key
[Search] - PEAK - NEXT PEAK LEFT
Description
MAX Specifies the maximum value search.
MIN Specifies the minimum value search.
TARG Specifies the target search. The target closest to the current marker
position
is searched for.
PEAK Specifies the peak search. The maximum peak (minimum peak if
the peak has been defined as negative peak) is
searched for.
OFF (initial value) Turns off the search function.