Chapter 16 309
GPIB Command Reference
E4TP
16. GPIB
Command Reference
( * , A - E )
E4TP
Syntax E4TP {OFF|M1|M2|APC7|PROBE}
E4TP?
Description Selects the adapter in the adapter setting.
Parameters
Query response {OFF|M1|M2|APC7|PROBE}<newline><^END>
Corresponding
key
[Cal] - ADAPTER [ ] - {NONE | 4TP 1M | 4TP 2M | 7mm 42942A | PROBE 42941A}
ECAL{P|A|B|C}
Syntax ECAL{P|A|B|C}
Description Measures the PHASE/OPEN/SHORT/LOAD data to calculate the setup data of the
selected adapter. The “ECALP” command measures the PHASE data; the “ECALA”
command, the OPEN data; the “ECALB” command, the SHORT data; the “ECALC”
command, LOAD data. These data must be measured for each adapter (selected with the
“E4TP” command on page 309). (No query)
Corresponding
key
[Cal] - ADAPTER [ ] - SETUP - {PHASE COMP [ ] | OPEN [ ] | SHORT [ ] | LOAD [ ]}
ECALDON
Syntax ECALDON
Description Finishes the measurement of the data for the adapter setup and stores the measured data
into the non-volatile memory. The setup data is stored as the data for the adapter selection
specified at the measurement (data is stored for each adapter selection). (No query)
Corresponding
key
[Cal] - ADAPTER [ ] - SETUP - done
Description
OFF (initial value) Specifies no adapter (NONE).
M1 Specifies the Agilent 16048G (4TP 1M).
M2 Specifies the Agilent 16048H (4TP 2M).
APC7 Specifies the Agilent 42942A (7mm 42942A).
PROBE Specifies the Agilent 42941A (PROBE 42941A).