292 Chapter 16
GPIB Command Reference
DCOMLOAD{R|L}
DCOMLOAD{R|L}
Syntax DCOMLOADR <numeric>[OHM]
DCOMLOADL <numeric>[H]
DCOMLOAD{R|L}?
Description For fixture compensation data measurement (“COM{A|B|C}” command on page 281), sets
the resistance value (“DCOMLOADR” command) and the inductance value
(“DCOMLOADL” command) of the LOAD standard used to measure LOAD
compensation data.
Parameters
If the specified parameter is out of the allowable setting range, the minimum value (if the
lower limit of the range is not reached) or the maximum value (if the upper limit of the
range is exceeded) is set.
Query response {numeric}<newline><^END>
Corresponding
key
[Cal] - FIXTURE COMPEN - DEFINE VALUE - {LOAD RESIST (R) | LOAD INDUCT (L)}
<numeric>
Description Resistance value or inductance value
Range For resistance value: -100E9 to 100E9
For inductance value: -1E6 to 1E6
Initial value For resistance value: 50
For inductance value: 0
Unit For resistance value: Ω (ohm)
For inductance value: H (henry)
Resolution 1E-12
Page view 291