Chapter 16 279
GPIB Command Reference
CLEL
16. GPIB
Command Reference
( * , A - E )
CLEL
Syntax CLEL
Description In the creation/edit of the list sweep table, deletes all segments. Executing this command
during the execution of segment creation/edit causes an error. If list sweep table
creation/edit has not been started, executing this command automatically starts table
creation/edit and then deletes all segments. (No query)
NOTE Executing the “EDITDONE” command on page 310 brings the created/edited list sweep
table to take effect.
Corresponding
key
[Sweep] - EDIT LIST - CLEAR LIST
CLES
Syntax CLES
Description Clears the error queue, Status Byte Register, Operation Status Register, Standard Event
Status Register, and Instrument Event Status Register. This command has the same
function as the
“*CLS” command on page 258. (No query)
Corresponding
key
No front panel key is available to execute this function.