Appendix D 495
Error messages
Error number: 83
1. Chapter Title
2. Chapter Title D. Error messages
4. Chapter Title
5. Chapter Title
When an active controller was on the same GPIB bus, you attempted to set the 4294A as a
system controller. Unless another active controller is removed from the same GPIB bus,
you cannot set the 4294A as a system controller.
83 NOT VALID FOR COMPLEX MEASUREMENT
When the COMPLEX Z-Y measurement parameter (that is, complex parameter
measurement) is selected, a command to execute one of the following is sent. As a result,
the command is ignored.
Setting or reading an offset value in the trace calculation function (DATOVAL
command)
Assigning the main marker to the offset value in the trace calculation function
(MKROFS command)
Selecting DELTA % in the trace calculation function (DISP DELP command)
When making the above settings, first select a measurement parameter other than
COMPLEX Z-Y.
84 NOT VALID FOR SCALAR MEASUREMENT
When a measurement parameter other than COMPLEX Z-Y (that is, scalar parameter
measurement) is selected, a command to execute one of the following is sent. As a result,
the command is ignored.
Setting the secondary measurement parameter value of the fixed Δ marker
(DMKRAUV command)
Selecting DATA/MEM in the trace calculation function (DISP DDVM command)
When making the above settings, first select the COMPLEX Z-Y measurement parameter.
85 FORMAT NOT VALID FOR COMPLEX MEASUREMENT
When COMPLEX Z-Y (i.e., complex parameter measurement) was selected as a
measurement parameter, you attempted to select a display format not available for the
measurement parameter, for example, the linear scale format (FMT LINY command) and,
as a result, the command was ignored. When COMPLEX Z-Y is selected as a measurement
parameter, set the display format to the complex plane display format (command: FMT
COMP) or the polar coordinates display format (command: FMT
POLA).
86 FORMAT NOT VALID FOR SCALOR MEASUREMENT
When a measurement parameter other than COMPLEX Z-Y (i.e., scalar parameter
measurement) was selected, you attempted to select a display format not available for the
measurement parameter, for example, the complex plane display format (command: FMT
COMP) and, as a result, the command was ignored. When a measurement parameter other
than COMPLEX Z-Y is selected, select a display format from the linear scale format
(command: FMT LINY), the log scale format (command: FMT LOGY), or the percent
format (command: FMT PERC).
87 THETA MEASUREMENT NOT SELECTED
When the setting of the phase (q) was not active as a measurement parameter, you
attempted to execute a command to set the phase unit (PHAU DEG or PHAU RAD) or a
command to set the phase expansion display (EXPP ON or EXPP OFF) and, as a result, the
command was ignored. The active status of the phase (θ) setting means:
•|Z|-θ or |Y|-θ is selected as a measurement parameter and trace B is active.