268 Chapter 16
GPIB Command Reference
ANARFULL
ANARFULL
Syntax ANARFULL
Description Sets the analysis range to the entire sweep range when performing waveform analysis
using the waveform analysis command. (No query)
Corresponding
key
No front panel key is available to execute this function.
ANASEGM
Syntax ANASEGM <numeric>
ANASEGM?
Description When the sweep type is set to the list sweep, selects, from the list sweep table, the
waveform analysis target segment when performing analysis using the waveform analysis
function. The waveform analysis target segment number is initialized when the sweep type
is set to the list sweep from other than the list sweep. Therefore, you have to set the
waveform analysis target segment number each time.
Parameters
If the specified parameter is out of the allowable setting range, the minimum value (if the
lower limit of the range is not reached) or the maximum value (if the upper limit of the
range is exceeded) is set.
Query response {numeric}<newline><^END>
If the sweep type is not the list sweep, 0 is always read out.
Corresponding
key
No front panel key is available to execute this function.
AUTO
Syntax AUTO
Description Automatically sets scale parameters so that the trace fits on the screen (executes the auto
scale). (No query)
Corresponding
key
[Scale Ref] - AUTO SCALE
<numeric>
Description Waveform analysis target segment number
Range 1 to the number of segments in the list sweep table
Initial value 1
Resolution 1