276 Chapter 16
GPIB Command Reference
CALST
CALST
Syntax CALST {ON|OFF|1|0}
CALST?
Description Toggles on and off the user calibration function. If the user calibration data is not stored
(executed with the
“CALDON” command on page 274), executing this command to turn
on the calibration causes an error and the command is ignored.
Parameters
Query response {1|0}<newline><^END>
Corresponding
key
[Cal] - USER CAL - CORRECTION on OFF
CBRI
Syntax CBRI <numeric>[PCT]
CBRI?
Description Sets the brightness of the display color of the item selected with the “COLO” command on
page 280.
Parameters
Query response {numeric}<newline><^END>
Corresponding
key
[Display] - more 1/2 - MODIFY COLOR - MODIFY ITEM COLOR - MODIFY COLOR -
BRIGHTNESS
Description
ON or 1 Turns on the user calibration function.
OFF or 0 (initial value) Turns off the user calibration function.
<numeric>
Description Percentage of brightness (the brightest state: 100 %)
Range 0 to 100
Initial value Varies depending on the display item selected with the “COLO”
command on page 280.
Unit % (percentage)
Resolution 1