Chapter 16 451
GPIB Command Reference
TARR?
16. GPIB
Command Reference
( S - Z , :PROG )
TARR?
Syntax TARR? <numeric 1>
Description Within the waveform analysis range specified with the “ANARANG” command on page
267, searches the analysis range from left to right for a point of the specified measurement
value, and reads out the sweep parameter value at the first detected point. If the specified
measurement parameter value is not detected, 0 is read out.(Query only)
Parameters
If the specified parameter is out of the setting range, an error occurs and the command is
ignored.
Query response {numeric 2}<newline><^END>
Corresponding
key
No front panel key is available to execute this function.
<numeric 1>
Description Measurement parameter value you want to search for
Range -9.9E37 to 9.9E37
Unit Varies depending on the measurement parameter (refer to the
explanation of unit in
“SCAL” on page 423).
{numeric 2}
Description Sweep parameter value of the detected point
Unit Varies depending on the sweep parameter (refer to the explanation
of unit in
“CENT” on page 277).