Agilent Technologies 4294A Service Manual Page 363

  • Download
  • Add to my manuals
  • Print
  • Page
    / 518
  • Table of contents
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 362
Chapter 16 363
GPIB Command Reference
NEXP
16. GPIB
Command Reference
( O - R )
NEXP
Syntax NEXP
Description When the measurement results are listed as a result of the execution of the “LISV”
command on page 341, displays the next page of the list. To display the previous page,
execute the “PREP” command on page 404. (No query)
Corresponding
key
[Copy] - SELECT CONTENTS - LIST VALUE - NEXT PAGE
NEXPK?
Syntax NEXPK?
Description Within the waveform analysis range set with the “ANARANG” command on page 267,
searches for the maximum peak having a value less than the value of the peak that has been
detected at the last execution of this command or the
“PEAK?” command on page 397, and
reads out the measurement parameter value and the sweep parameter value of the peak. If
there are several applicable peaks, reads out the value of the leftmost peak within the
analysis range. If no peak exists, 0 is read out.
The peak searched for at the execution of this command satisfies the condition set with the
“THRR” command on page 452. It is not related to the peak definition in the marker search
function. (Query only)
Query response {numeric 1},{numeric 2}<newline><^END>
Readout value are as follows:
{numeric 1}: Measurement parameter value of the detected peak
{numeric 2}: Sweep parameter value of the detected peak
Corresponding
key
No front panel key is available to execute this function.
Page view 362
1 2 ... 358 359 360 361 362 363 364 365 366 367 368 ... 517 518

Comments to this Manuals