338 Chapter 16
GPIB Command Reference
LIMUSTOP
LIMUSTOP
Syntax LIMUSTOP <numeric>[DEG|RAD|OHM|SIE|H|F|PCT]
LIMUSTOP?
Description In the creation/edit of the limit line table, sets the upper limit of the stop point of each
segment (specified with the
“LIMSTOP” command on page 336). Use this command,
together with the “LIMLSTOP” command on page 329, when specifying the limits using
the upper limit and the lower limit.
If segment creation/edit has not been started, executing this command automatically starts
the edit of the edit target segment and then sets the upper limit of the stop point.
If the creation/edit of the limit line table has not been started, executing this command
automatically starts table creation/edit, starts the edit of the edit target segment, and then
sets the upper limit of the stop point.
Parameters
If the specified parameter is out of the allowable setting range, the minimum value (if the
lower limit of the range is not reached) or the maximum value (if the upper limit of the
range is exceeded) is set.
Query response {numeric}<newline><^END>
Corresponding
key
[System] - LIMIT TEST - EDIT LIMIT LINE - EDIT - more 1/3 - UPPER LIMIT
<numeric>
Description Upper limit of the segment stop point
Range -1E9 to 1E9
Initial value When creating segment 1: 0
When creating an additional segment: Set value of the previous
segment
Unit Varies depending on the measurement parameter (refer to the
explanation of unit in
“SCAL” on page 423).