Chapter 16 435
GPIB Command Reference
SEGM
16. GPIB
Command Reference
( S - Z , :PROG )
SEGM
Syntax SEGM <numeric>
SEGM?
Description Sets the execution target segment used when no segment is specified at the execution of the
“SADD” command on page 415, “SDEL” command on page 425 and “SEDI” command
on page 434. The setting of the execution target segment, when the creation/edit of the list
sweep table is finished, is initialized.
If the creation/edit of the list sweep table is not started, executing this command
automatically starts table creation/edit and then sets the execution target segment.
Parameters
If the specified parameter is out of the allowable setting range, the minimum value (if the
lower limit of the range is not reached) or the maximum value (if the upper limit of the
range is exceeded) is set.
Query response {numeric}<newline><^END>
If the creation/edit of the list sweep table is not started, 0 is always read out.
Corresponding
key
[Sweep] - EDIT LIST - SEGMENT
<numeric>
Description Segment number you want to specify as the edit target
Range 1 to the number of segments in the list sweep table
Initial value Segment number of the last line in the list sweep table
Resolution 1