376 Chapter 16
GPIB Command Reference
OUTPDC?
OUTPDC?
Syntax OUTPDC?
Description In the dc bias level monitor function, if monitoring voltage or current is enabled (VOLT or
CURR has been specified with the
“BMON” command on page 272), reads out the dc bias
level voltage or current monitor value at all measurement points. (Query only)
Query response {numeric 1},{numeric 2},..,{numeric NOP}<newline><^END>
Where, NOP is the number of points.
If the oscillator level monitor function is set to off, the readout are invalid.
Corresponding
key
No front panel key is available to execute this function.
OUTPDCP?
Syntax OUTPDCP? <numeric 1>
Description In the dc bias level monitor function, if monitoring voltage or current is enabled (VOLT or
CURR has been specified with the
“BMON” command on page 272), reads out the dc bias
level voltage or current monitor value of the specified measurement point. (Query only)
Parameters
If the specified parameter is out of the allowable setting range, the minimum value (if the
lower limit of the range is not reached) or the maximum value (if the upper limit of the
range is exceeded) is set.
Query response {numeric 2}<newline><^END>
If the oscillator level monitor function is set to off, the readout is invalid.
Corresponding
key
No front panel key is available to execute this function.
<numeric 1>
Description Number of the measurement point you want to read out (the
number assigned in order from the left edge assuming that the
number of the leftmost measurement point is 1)
Range 1 to the number of points
Resolution 1
{numeric 2}
Description Voltage or current monitor value of the dc bias level