Agilent Technologies 4294A Service Manual Page 352

  • Download
  • Add to my manuals
  • Print
  • Page
    / 518
  • Table of contents
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 351
352 Chapter 16
GPIB Command Reference
MKRCOUP
MKRCOUP
Syntax MKRCOUP {ON|OFF|1|0}
MKRCOUP?
Description As the marker move setting, selects the mode in which the markers on traces A and B are
coupled when they are moved (coupled marker mode) or the mode in which the markers on
traces A and B are moved separately. (uncoupled marker mode).
If the markers on traces A and B are at different positions when changing the mode from
the uncoupled marker mode to the coupled marker mode, to the position of the marker on
the active trace (set with the
“TRAC” command on page 457), the other marker is moved.
Parameters
Query response {1|0}<newline><^END>
Corresponding
key
[Marker] - MKR COUPLE ON off
MKRDSPAN
Syntax MKRDSPAN
Description For the active trace (set with the “TRAC” command on page 457), sets the sweep
parameter value difference between the Δ marker and the marker as the span value of the
sweep range. If the Δ marker mode is off, executing this command causes an error and the
command is ignored. (No query)
Corresponding
key
[Marker] - MKRΔ → SPAN
MKRL
Syntax MKRL {ON|OFF|1|0}
MKRL?
Description For the active trace (set with the “TRAC” command on page 457), toggles on and off the
marker list function. If the marker function is off, executing this command causes an error
and the command is ignored.
Parameters
Query response {1|0}<newline><^END>
Corresponding
key
[Utility] - MKR LIST on OFF
Description
ON or 1 (initial value) Specifies the coupled marker mode.
OFF or 0 Specifies the uncoupled marker mode.
Description
ON or 1 Turns on the marker list function.
OFF or 0 (initial value) Turns off the marker list function.
Page view 351
1 2 ... 347 348 349 350 351 352 353 354 355 356 357 ... 517 518

Comments to this Manuals

No comments